Model Number 0998-00-3023-55 |
Device Problem
Low Readings (2460)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 11/09/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
A getinge field service engineer (fse) was dispatched to evaluate this unit.The fse checked the iabp with a trainer for more than one hour without issue.The fse performed all required functional tests.The iabp unit was cleared for clinical use, and released to the customer.
|
|
Event Description
|
It was reported that the cs300 intra-aortic balloon pump (iabp) augmented pressure was lower than the systolic pressure.It is unknown the circumstances under which the event occurred.It is also unknown if there was patient involvement.However, there was no adverse event reported.
|
|
Manufacturer Narrative
|
Historical data analysis: (4109/213) the review of the historical data indicates that no other similar complaint was reported for the same lot/serial number and reported failure mode.Trend analysis: (4110/213) the overall 12 month product complaint trend data for the period ( dec-2019 to nov-2020 ) was reviewed.There were no triggers identified for the review period.
|
|