MEDTRONIC PUERTO RICO OPERATIONS CO. INTELLIS; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF Back to Search Results
Model Number 97715
Device Problems High impedance (1291); Shipping Damage or Problem (1570)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/17/2020
Event Type  malfunction  
Manufacturer Narrative
Concomitant medical products: product id: 977a260, serial#: (b)(4), implanted: (b)(6) 2020, product type: lead.Product id: 977a260, serial# (b)(4), implanted: (b)(6) 2020, product type: lead.Other relevant device(s) are: product id: 977a260, serial/lot #: (b)(4), ubd: 08-sep-2024, udi#: (b)(4).Product id: 977a260, serial/lot #: (b0(4), ubd: 08-sep-2024, udi#: (b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was reported that high impedance was observed on second electrode from channel port.There were no known factors related to this.Several impedance checks, switching channels, wiping off and cleaning several times.High impedance followed the electrode to each channel.The rep was unable to resolve the impedance issue, but the lead was implanted successfully.The issue was deemed resolved at the time of the report.Additional information was received from the manufacturer representative (rep) reporting that the cause of the high impedance issue was not determined.It was indicated that post-op programming was able to cover the patient¿s pain without using the electrodes with high impedances.It was indicated that the provided information had been confirmed with the physician.No further complications were reported/anticipated.Additional information was received.It was reported the high impedance was >40,000 ohms.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLIS
Type of Device
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
david gustafson
7000 central avenue ne rcw215
minneapolis, MN 55432
7635149628
MDR Report Key10929814
MDR Text Key219378410
Report Number3004209178-2020-20934
Device Sequence Number1
Product Code LGW
UDI-Device Identifier00643169781702
UDI-Public00643169781702
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P840001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 12/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/01/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/28/2021
Device Model Number97715
Device Catalogue Number97715
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/25/2020
Date Device Manufactured08/31/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age77 YR
-
-