• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number N/A
Device Problems No Audible Prompt/Feedback (2282); Device Contamination with Body Fluid (2317)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/24/2020
Event Type  malfunction  
Manufacturer Narrative
A getinge service territory manager (stm) replaced the pneumatic module assembly then completed scheduled preventative maintenance (pm).All safety, functionality, and calibration checks were performed, and all tests passed to factory specifications.The iabp unit was cleared for clinical use, and released to the customer.
 
Event Description
It was reported that during use on a patient, the cardiosave intra-aortic balloon pump (iabp) experienced a blood back event with no alarm.It is unknown if there was any patient harm/injury; however there was no adverse event reported.
 
Event Description
It was reported that during use on a patient, the cardiosave intra-aortic balloon pump (iabp) experienced a blood back event with no alarm.It is unknown if there was any patient harm/injury; however there was no adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID TYPE B PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ 07430
MDR Report Key10931312
MDR Text Key219342054
Report Number2249723-2020-02047
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108391
UDI-Public10607567108391
Combination Product (y/n)N
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 12/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/02/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number0998-00-0800-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Date Manufacturer Received12/10/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-