OLYMPUS MEDICAL SYSTEMS CORP. EVIS EXERA II VIDEO SYSTEM CENTER Back to Search Results
Model Number CV-180
Device Problem No Display/Image (1183)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  Malfunction  
Manufacturer Narrative

The suspect device was returned to olympus and evaluated. The reported problem was confirmed. A faulty rf board was found, causing no image on the sdi output signals. When capturing with the sdi signal, it would disable the unit.

 
Event Description

A user facility reported to olympus that the usdi port was dropping the signal. There was no patient injury or harm, associated with the problem, reported to olympus.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameEVIS EXERA II VIDEO SYSTEM CENTER
Type of DeviceVIDEO SYSTEM CENTER
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA 192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key10931943
MDR Text Key227214372
Report Number8010047-2020-09681
Device Sequence Number1
Product Code FAJ
Combination Product (Y/N)N
Reporter Country CodeUS
PMA/PMN NumberK133538
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type USER FACILITY
Reporter Occupation
Type of Report Initial,Followup
Report Date 01/13/2021
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received12/02/2020
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device MODEL NumberCV-180
Was Device Available For Evaluation? Device Returned To Manufacturer
Date Returned to Manufacturer11/02/2020
Is The Reporter A Health Professional? No
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received01/04/2021
Was Device Evaluated By Manufacturer? Yes
Date Device Manufactured10/03/2006
Is The Device Single Use? No
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-