MAUDE Adverse Event Report: MEDOS INTERNATIONAL SàRL CH SINGLE-INNER SETSCREW; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DDD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SàRL CH SINGLE-INNER SETSCREW; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DDD Back to Search Results
Model Number 179702000
Device Problem Device Slipped (1584)
Patient Problem Failure of Implant (1924)
Event Type  Injury  
Manufacturer Narrative
Additional procode: kwp;kwq;mnh;mni;osh.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.(b)(6).Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Device report from synthes reports an event in (b)(6) as follows: it was reported that on an unknown date six (6) set screws were loose in the patient ten (10) dats after scoliosis surgery.There is no further information available.This report is for one (1) single-inner setscrew.This is report 1 of 6 for (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h4, h6- the single inner setscrews (p/n: 179702000, lot number: 278172) was received at us cq.Upon visual inspection, the device was found intact.The provided x-ray show the set screw was migrated.The complaint condition was confirmed as implant was shifted after implantation.There is no indication that a design or manufacturing issue has caused the complaint condition.It is possible that the the implant was not tightened properly during implantation.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Device history lot =he dhr of product code 179702000, lot 278172, was reviewed and no non-conformance was observed during the manufacturing process.The product was released on april 6, 2020.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported the revision procedure occurred on (b)(6) 2020.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE-INNER SETSCREW
Type of Device
ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DDD
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL CH
chemin-blanc 38
le locle 02400
SZ  02400
MDR Report Key10934764
MDR Text Key219159117
Report Number1526439-2020-02326
Device Sequence Number1
Product Code NKB
UDI-Device Identifier10705034064307
UDI-Public(01)10705034064307
Combination Product (y/n)N
PMA/PMN Number
K160904
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 11/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/02/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number179702000
Device Catalogue Number179702000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/07/2020
Date Manufacturer Received01/17/2021
Patient Sequence Number1
Treatment
SINGLE-INNER SETSCREW; SINGLE-INNER SETSCREW; SINGLE-INNER SETSCREW; SINGLE-INNER SETSCREW; SINGLE-INNER SETSCREW; SINGLE-INNER SETSCREW; SINGLE-INNER SETSCREW; SINGLE-INNER SETSCREW; SINGLE-INNER SETSCREW; SINGLE-INNER SETSCREW
Patient Outcome(s) Required Intervention;
-
-