Catalog Number 8065752015
Device Problem Therapeutic or Diagnostic Output Failure (3023)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/08/2020
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A surgeon reported the aspiration became weak during a cataract procedure.The procedure was completed after replacing the product with another one.There was no harm to the patient.
 
Manufacturer Narrative
A review of the device history record traceable to the reported lot number, indicates that the product was processed and released according to the product¿s acceptance criteria the returned sample was visually inspected and no obvious defects were found.The administration manifold and the infusion cannula line were not returned.The non-invasive flow sensor (nifs) on the cassette housing was in good condition.A calibrated console representing the current software version was used to test the sample.The ball in the drip chamber¿s check valve moved freely per specification.The light-emitting diode (led) rings on the console turned green as the probe connectors were engaged to the console, indicating the proper communication between the probe and the console.The sample could prime and pass intraocular pressure (iop) calibration successfully, with lab stock administration manifold and the infusion cannula.No anomalies were observed during priming.No system message code was generated during testing.Fluid flowed from the balanced salt solution (bss) bottle to the drain bag without any interfering.No leakage was detected from the pump elastomer or on the pump area of the fluidics module.Cleaning process was able to be performed after functional test had completed.The root cause of the customer's complaint could not be established.The returned sample functioned per specifications.After an investigation of this complaint, it has been determined, that this sample functioned per specifications.Therefore, no corrective action is required at this time.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION SURGICAL PROCEDURE PAK
Type of Device
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
MDR Report Key10937144
MDR Text Key219519920
Report Number1644019-2020-00647
Device Sequence Number1
Product Code LRO
Combination Product (y/n)N
PMA/PMN Number
K880961
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 03/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/02/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2021
Device Catalogue Number8065752015
Device Lot Number2312364H
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/17/2020
Date Manufacturer Received02/22/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
CONSTELLATION VISION SYSTEM
-
-