Model Number HX-202UR |
Device Problem
Device Fell (4014)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 11/04/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
As the device has not been returned for evaluation, we are unable to determine a root cause for the reported event.If additional information becomes available or the device is returned for evaluation, a supplemental report will be filed.
|
|
Event Description
|
A user facility reported that a single use repositionable clip fell off during a colonoscopy with polypectomy procedure.There was no patient injury due to the reported issue.The clip was retrieved and another one was used to complete the procedure.No additional information has been obtained.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the legal manufacturer's investigation, the phenomenon likely occurred because the clip was not grasping the tissue when the clip was placed.Therefore, the clip was released from the distal end of the insertion portion and detached.
|
|