Model Number HX-202UR
Device Problem Device Fell (4014)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/04/2020
Event Type  malfunction  
Manufacturer Narrative
As the device has not been returned for evaluation, we are unable to determine a root cause for the reported event.If additional information becomes available or the device is returned for evaluation, a supplemental report will be filed.
 
Event Description
A user facility reported that a single use repositionable clip fell off during a colonoscopy with polypectomy procedure.There was no patient injury due to the reported issue.The clip was retrieved and another one was used to complete the procedure.No additional information has been obtained.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the legal manufacturer's investigation, the phenomenon likely occurred because the clip was not grasping the tissue when the clip was placed.Therefore, the clip was released from the distal end of the insertion portion and detached.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE USE REPOSITIONABLE CLIP
Type of Device
SINGLE USE REPOSITIONAL CLIP
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key10938493
MDR Text Key219369662
Report Number8010047-2020-09752
Device Sequence Number1
Product Code PKL
UDI-Device Identifier04953170385940
UDI-Public04953170385940
Combination Product (y/n)N
PMA/PMN Number
K123601
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 02/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/03/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHX-202UR
Device Lot Number9XK
Was the Report Sent to FDA? No
Date Manufacturer Received02/02/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-