MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. OES CYSTONEPHROFIBERSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. OES CYSTONEPHROFIBERSCOPE Back to Search Results
Model Number CYF-5
Device Problem Microbial Contamination of Device (2303)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/05/2020
Event Type  Malfunction  
Manufacturer Narrative

The subject device has not been returned to omsc but was returned to olympus (b)(4). (ofr). Ofr sent the device to a third party laboratory for microbiological testing. As a result of the testing, no microbe was detected from the sample collected from the all channels of the device. The testing result cleared the french guideline. The exact cause of the reported event could not be conclusively determined at this time. If additional information becomes available, this report will be supplemented.

 
Event Description

Olympus medical systems corp. (omsc) was informed that as a result of multiple microbiological testing by the user facility, following microbes were detected from the sample collected from the subject device. [first time; (b)(6) 2020] instrument channel: bacillus megaterium (2 cfu) [second time; (b)(6) 2020] instrument channel: bacillus simplex (1 cfu) other detailed information such as the reprocessing method was not provided. There was no report of infection associated with this report.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameOES CYSTONEPHROFIBERSCOPE
Type of DeviceCYSTONEPHROFIBERSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA 192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key10948838
MDR Text Key226708596
Report Number8010047-2020-09851
Device Sequence Number1
Product Code FAJ
Combination Product (Y/N)N
Reporter Country CodeFR
PMA/PMN NumberK032092
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type COMPANY REPRESENTATIVE,FOREIG
Reporter Occupation
Type of Report Initial,Followup
Report Date 01/15/2021
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received12/04/2020
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device MODEL NumberCYF-5
Was Device Available For Evaluation? No
Is The Reporter A Health Professional? No
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received01/07/2021
Was Device Evaluated By Manufacturer? Device Not Returned To Manufacturer
Date Device Manufactured02/14/2017
Is The Device Single Use? No
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-