Model Number 466FXXXX
Device Problems Failure to Align (2522); Unintended Movement (3026)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/05/2020
Event Type  Injury  
Manufacturer Narrative
It was reported that a patient underwent placement of an optease vena cava filter.The information provided indicated that the filter subsequently malfunctioned and caused filter migration, and filter tilt.The indication for the filter implant, procedural details and medical history of the patient have not been provided.There is currently no additional information available for review.The product was not returned for analysis and the sterile lot number has not been provided; therefore, no device analysis nor device history record review could be performed.The optease vena cava filter is indicated for use in the prevention of recurrent pulmonary embolism (pe) via percutaneous placement in the vena cava for patients in which anticoagulants are contraindicated, anticoagulant therapy for thromboembolic disease has failed, emergency treatment following massive pulmonary embolism where anticipated benefits of conventional therapy are reduced or for chronic, recurrent pulmonary embolism where anticoagulant therapy has failed, or is contraindicated.The purpose of a vena cava filter is to catch thrombus from the lower extremities as it travels along normal blood flow patterns up towards the heart.Post implant imaging has not been provided.Ivc filter tilt has been associated with operator technique, vessel characteristics, specifically asymmetry and tortuosity.Inferior vena cava (ivc) filter migration is a known potential adverse event associated with all ivc filter implants and is listed in the instructions for use (ifu) as such.Possible causes for filter migration include mega cava, wire entrapment during central venous catheter placement, ¿sail¿ effect (cranial migration) of large clot burden within the filter, mechanical device failure, and operator error.Physiologic causes of migration may result from temporary dysmorphism of the inferior vena cava including bending, coughing or valsalva maneuvers resulting in dislodgment of the filter.Some studies suggest that strenuous physical activity and increased intra-abdominal pressure can lead to migration of ivc filters.Without the procedural films or post-placement imaging and the limited information provided, the report of migration and tilt could not be confirmed or further clarified.Given the limited information currently available for review, there is nothing to suggest that the reported events are related to the design and manufacturing process of the device; therefore, no corrective action will be taken.Should additional information become available, the file will be updated accordingly.
 
Event Description
As reported by the legal team, the patient underwent placement of the optease vena cava filter.The filter subsequently malfunctioned and caused injury and damages to the patient, including, but not limited to, physical and emotional damages from tilt, migration and the resultant symptoms.As a direct and proximate result of these malfunctions, the patient suffered life-threatening injuries and damages, and required extensive medical care and treatment.As a further proximate result, the patient has suffered and will continue to suffer significant medical expenses, and pain and suffering, and other damages.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
466FXXXX
Type of Device
FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
CORDIS CASHEL
cahir road
cashel, tipperary 0000
EI  0000
Manufacturer (Section G)
CORDIS CASHEL
cahir road
cashel, tipperary 0000
EI   0000
Manufacturer Contact
karla castro
14201 nw 60th ave
miami lakes, FL 33014
7863138372
MDR Report Key10949928
MDR Text Key219868155
Report Number1016427-2020-04605
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/04/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number466FXXXX
Device Catalogue Number466FXXXX
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/11/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-