• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GREINER BIO-ONE NA INC. VACUETTE BLOOD COLLECTION TUBES EDTA; EVACUATED BLOOD COLLECTION TUBE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GREINER BIO-ONE NA INC. VACUETTE BLOOD COLLECTION TUBES EDTA; EVACUATED BLOOD COLLECTION TUBE Back to Search Results
Device Problem Therapeutic or Diagnostic Output Failure (3023)
Patient Problem Insufficient Information (4580)
Event Type  malfunction  
Manufacturer Narrative
Complaint (b)(4).No samples received for evaluation.No pictures received.No material # nor batch # was provided by the customer.No clarification or further information was received from the customer.Unfortunately, without basic information, a thorough investigation is not possible.This complaint is closed as cannot be determined due to insufficient information.
 
Event Description
Customer states increase in clumped platelets and clotted samples since switching to greiner lavender top tubes, especially with outreach work.When reviewing the slide (peripheral blood smear) they see micro-clumps,.2, 3-4 platelets sticking together.The instrument does not flag these but (they) will go and perform a fluorescent platelet under the microscope if needed.Not all sites have this option to perform fluorescent platelet.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VACUETTE BLOOD COLLECTION TUBES EDTA
Type of Device
EVACUATED BLOOD COLLECTION TUBE
Manufacturer (Section D)
GREINER BIO-ONE NA INC.
4238 capital drive
monroe NC 28110
Manufacturer (Section G)
GREINER BIO-ONE NA INC.
4238 capital drive
monroe NC 28110
Manufacturer Contact
manfred abel
4238 capital drive
monroe, NC 28110
7042617823
MDR Report Key10951725
MDR Text Key229077839
Report Number1125230-2020-00067
Device Sequence Number1
Product Code JKA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K042927
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/05/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/09/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-