Model Number MAJ-891
Device Problems Break (1069); Contamination /Decontamination Problem (2895)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
The device was not returned to olympus medical systems corp. (omsc), but returned to olympus repair center for the evaluation. The evaluation of the device confirmed the followings; component analysis showed that the brown foreign matter was rust. Omsc guessed that the screw cover coming off was caused by deterioration of the adhesive to secure the screw cover due to chemical or physical stress. The rust may have been caused by corrosion of residues such as chemicals. If significant additional information is received, this report will be supplemented.
 
Event Description
A user reported that some the parts of the device came off and the lever got loose during reprocess. Olympus inspected the device at the service department of olympus medical systems corp. (omsc) and found that the screw cover had come off. It was also found that brown foreign matters adhered between the stop cock and the main body. There was no report of patient injury associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameFORCEPS/IRRIGATION PLUG (ISOLATED TYPE)
Type of DeviceFORCEPS/ IRRIGATION PLUG
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA 192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key10956810
MDR Text Key219875071
Report Number8010047-2020-09909
Device Sequence Number1
Product Code FGB
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K912120
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 12/07/2020
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received12/07/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberMAJ-891
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/16/2020
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received11/13/2020
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-