• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER CENTURION VISION SYSTEM; UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER CENTURION VISION SYSTEM; UNIT, PHACOFRAGMENTATION Back to Search Results
Model Number ASKU
Device Problem Improper or Incorrect Procedure or Method (2017)
Patient Problem Endophthalmitis (1835)
Event Type  Injury  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A surgeon reported a patient experienced endophthalmitis following an ophthalmic surgical procedure.A company representative has visited the site.It was noted that the ophthalmic irrigating solution bag was "hit" with a syringe to inject epinephrine, then hit again to connect the tubing of the cassette, but there was no disinfection with an alcohol wipe.This step has disappeared from their practice.Sterilization processes are called into question for the system and handpieces.Additional information has been requested but not yet received.This is one of two reports from this facility.
 
Manufacturer Narrative
The customer did not request service for the system.The system manufacturing device history record (dhr) was reviewed.Based on assessment, the product met specifications at the time of release.With no additional, related information provided, the customer reported events could not be confirmed.The root cause of the reported event cannot be determined conclusively.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTURION VISION SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
MDR Report Key10960175
MDR Text Key219995680
Report Number2028159-2020-01065
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
PMA/PMN Number
K121555
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 04/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/07/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberASKU
Device Catalogue NumberASKU
Device Lot NumberASKU
Was Device Available for Evaluation? No
Date Manufacturer Received03/26/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-