The device was returned for analysis.The reported material separations were able to be confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents from this lot.Based on the condition of the returned device being consistent with preparation and advancement onto the guide wire, it is possible the reported/noted device damages occurred during use; however due to the limited amount of case details the investigation determined a conclusive cause for the reported/noted device damages cannot be determined.There is no indication of a product quality issue with respect to manufacture, design or labeling.Na.
|