OLYMPUS MEDICAL SYSTEMS CORP. URETERO-RENO VIDEOSCOPE Back to Search Results
Model Number URF-V
Device Problem Mechanical Jam (2983)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
The device was returned to olympus for evaluation. The repair center confirmed the bending manipulation and insertion section defects caused deterioration of the angulation mechanism. This deterioration of the angulation mechanism was due to rust. The channel was scraped and leaking, fluid invaded the scope through the leak and caused damaged to the angulation system. The equipment was repaired and returned to the customer. A supplemental will be submitted if additional information becomes available following investigation.
 
Event Description
The device was returned to the olympus repair center by the customer. During evaluation, the repair center found bending manipulation and insertion section defects. There was no patient involvement; the reportable malfunction was found during service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameURETERO-RENO VIDEOSCOPE
Type of DeviceURETERO-RENO VIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA 192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key10971641
MDR Text Key227200982
Report Number8010047-2020-10056
Device Sequence Number1
Product Code FGB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K072957
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation
Type of Report Initial,Followup
Report Date 01/06/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received12/08/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberURF-V
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/03/2020
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received01/06/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-