Model Number PFSR110137
Device Problems Failure to Calibrate (2440); Mechanics Altered (2984)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/23/2020
Event Type  malfunction  
Event Description
It was reported that, during femur burring during a navio tka procedure, they got a handpiece exposure control motor error forcing them to recalibrate the handpiece.This happened three times (it would calibrate and home without issues but would immediately boot them back to the calibration screen every time they got to the burring screen).They decided to get a new handpiece and were able to finish the case without further problems.There was a delay of fewer than 30 minutes.No other complications were reported.
 
Manufacturer Narrative
H3, h6: the navio handpiece, part number pfsr110137, serial (b)(6) and intended to be used for treatment, was returned for evaluation.A relationship between the reported event and the device was established.The reported problem could not be visually confirmed.A functional evaluation was performed.The reported problem was confirmed.The "cannot communicate with the handpiece" error appeared before the test could start 3 times.A review of manufacturing records indicate the device met all specifications upon release into distribution.A complaint history review for similar reported/confirmed complaints has identified prior events.The most probable cause of the reported problem is internal electrical failure in cable.Although no further containment or corrective action is recommended or required at this time, all complaints are monitored and trended through post market surveillance activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAVIO HANDPIECE
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
MDR Report Key10972168
MDR Text Key220476408
Report Number3010266064-2020-02104
Device Sequence Number1
Product Code OLO
UDI-Device Identifier00885556628515
UDI-Public00885556628515
Combination Product (y/n)N
PMA/PMN Number
K191223
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 09/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/08/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberPFSR110137
Device Catalogue NumberPFSR110137
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/08/2020
Date Manufacturer Received09/27/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-