The devices were received and the complaint was confirmed.Examination revealed 10 out of 19 lock screws returned were found to have sever inferior surfaces damage.Review of the damage observed is consistent with rod interference during lock screw placement that can disallow complete full lock screw/ rod contact and secure final lock down and damage lock screws.No additional investigation required.Labeling review: ".Potential adverse events and complications: potential risks identified with the use of this system, which may require additional surgery, include: bending, fracture or loosening of implant component(s), loss of fixation.".".Warnings, cautions and precautions: if healing is delayed, or does not occur, the implant may eventually loosen.".".Step 4: rod & lock screw insertion after cutting the rod to length and contouring, place the rod into the implants and insert lock screws to provisionally secure the rod.1.The rod holder may be used to assist in placing the rod (fig.13).Use the longitudinal lines to ensure the rod is placed in proper sagittal alignment.To release the rod, depress the button at the center of the proximal ratchet.2.The multi-load lock screw starter can be used to deliver up to eight lock screws when no reduction is required (fig.14).O turn the gray central sleeve counterclockwise to fully retract the distal sleeve.O insert the distal end into the lock screw (5.5 or 6.0 mm, cannulated only), ensuring the silver side of the lock screw is facing up.O once the lock screws have been loaded, turn the central sleeve clockwise to compress the lock screws together.Do not hold onto the central sleeve during lock screw delivery, as this will cause the distal sleeve to retract, preventing adequate compression of the lock screws.O after delivery of each lock screw, rotate the central sleeve clockwise until resistance is felt prior to delivering the next lock screw.Tip: the lock screws on the multi-load lock screw starter must be compressed by the central sleeve after delivery of each lock screw to ensure proper engagement into the tulip.3.If reduction is required, use the lock screw starter to load a single lock screw and deliver down the reducer of choice.".".Post-operative warnings: damage to the weight-bearing structures can give rise to loosening of the components, dislocation and migration.".".Warnings, cautions and precautions: care should be taken to insure that all components are ideally fixated prior to closure.All implants should be used only with the appropriately designated instrument (reference surgical technique).".".All lock screws should be final-tightened with the counter-torque and torque t-handle.Do not final-tighten through compression instruments (e.G.C/d rack and figure 8 compressor) in the set, as the rod may not be able to normalize to the tulip.Be cautious not to over compress or distract as you can loosen the screws in the spine and potentially pull out the screw.The bulleted portion of the nose of the rod and the faceted portion of the rod (where the inserter locks down on the rod) must extend fully outside of the most inferior or most superior tulip on the construct.The set screw cannot be locked down on this unusable portion of the rod, as this may compromise the stability of the construct.All set screws should be final-tightened with the counter- torque and torque t-handle.Do not final-tighten through compression instruments in the set, as the rod may not be able to normalize to the tulip.".
|
The patient underwent a posterior fixation procedure with reline products on (b)(6) 2020.On (b)(6) 2020, a revision surgery was performed to replace a loosened lock screw of sai with new one.At that time, the physician noticed that the multiple others lock screw had been also loose.As a result, the physician replaced the 19 lock screws with new ones.
|