• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE Back to Search Results
Model Number V60 PLUS
Device Problem Therapeutic or Diagnostic Output Failure (3023)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6) 2020.Date of report: 08dec2020.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Event Description
It was reported to philips that the touchscreen was not responding in the lower right corner.The touchscreen calibration was performed several times, but did not solve the lower right corner.The device was not in clinical use at the time of the event.There was no report of patient or user harm.An international service engineer evaluated the device and confirmed that the touchscreen required replacement.The service engineer replaced the touchscreen and completed a full preventative maintenance.The device successfully passed all testing after repairs were completed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V60 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, INC
2271 cosmos court
carlsbad CA 92011
MDR Report Key10972791
MDR Text Key220916418
Report Number2031642-2020-04445
Device Sequence Number1
Product Code MNT
Combination Product (y/n)N
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Type of Report Initial
Report Date 10/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/08/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberV60 PLUS
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received10/28/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/10/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-