RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
|
Back to Search Results |
|
Model Number V60 PLUS |
Device Problem
Therapeutic or Diagnostic Output Failure (3023)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event: (b)(6) 2020.Date of report: 08dec2020.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
|
|
Event Description
|
It was reported to philips that the touchscreen was not responding in the lower right corner.The touchscreen calibration was performed several times, but did not solve the lower right corner.The device was not in clinical use at the time of the event.There was no report of patient or user harm.An international service engineer evaluated the device and confirmed that the touchscreen required replacement.The service engineer replaced the touchscreen and completed a full preventative maintenance.The device successfully passed all testing after repairs were completed.
|
|