• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEXCOM, INC. DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEXCOM, INC. DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 9500-64
Device Problem Imprecision (1307)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/20/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that an inaccuracy between the continuous glucose monitor (cgm) and the blood glucose (bg) meter occurred. The sensor was inserted into the abdomen. No data was provided for evaluation. Confirmation of the allegation and a probable cause could not be determined. The reported glucose values fall within the d zone of the parkes error grid. No injury or medical intervention was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameDEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM
Type of DeviceCONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence dr.
san diego CA 92121
Manufacturer Contact
ashley spoto
6340 sequence dr.
san diego, CA 92121
8582000200
MDR Report Key10980230
MDR Text Key220715467
Report Number3004753838-2020-152950
Device Sequence Number1
Product Code MDS
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P120005
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Reporter Occupation
Type of Report Initial
Report Date 12/09/2020
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received12/09/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number9500-64
Device Catalogue NumberSTS-JP-004
Was Device Available for Evaluation? No
Is the Reporter a Health Professional?
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received11/20/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-