Model Number V60
Device Problems Display or Visual Feedback Problem (1184); Failure to Shut Off (2939); Therapeutic or Diagnostic Output Failure (3023)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  Malfunction  
Manufacturer Narrative

Date of event: (b)(6) 2020. (b)(6) 2020. Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.

 
Event Description

The biomed called philips remote service engineer (rse) and reported being unable to shut the unit off. The biomed advised error codes consistent with alarm light emitting diode (led) failed, primary alarm failed, backup alarm failed and ventilator restarted due to anomalies detected during operation. Found in the significant event logs. The rse advised the biomed if have a spare power management (pm) printed circuit board assembly (pcba) or motor controller (mc) pcba to help diagnose. The biomed requested bench repair. The rse emailed the bench form. The device did not have patient involvement at the time the issue was discovered, therefore, there was no patient or user harm reported.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameV60 VENTILATOR
Type of DeviceVENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, INC
2271 cosmos court
carlsbad CA 92011
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS
carlsbad CA
Manufacturer Contact
bill cole
2271 cosmos court
carlsbad, CA 92011
9093746996
MDR Report Key10980379
MDR Text Key221272940
Report Number2031642-2020-04467
Device Sequence Number1
Product Code MNT
Combination Product (Y/N)N
Reporter Country CodeUS
PMA/PMN NumberK082660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type USER FACILITY
Reporter Occupation
Type of Report Initial
Report Date 11/12/2020
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received12/09/2020
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator NO INFORMATION
Device MODEL NumberV60
Device Catalogue Number1053614
Was Device Available For Evaluation? Yes
Is The Reporter A Health Professional?
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received11/12/2020
Was Device Evaluated By Manufacturer? Yes
Date Device Manufactured09/26/2018
Is The Device Single Use? No
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-