There was no patient involvement.Serial number is unknown.This information will be provided in a supplemental report if made available.As the serial number is unknown, the device manufacture date could not be determined.This information will be provided in a supplemental report if made available.Livanova (b)(4) implemented a field safety notice for disinfection and cleaning of heater-cooler devices.The z number is z-2076/2081-2015.Livanova (b)(4) manufactures the heater-cooler system 3t.The incident occurred in (b)(6).Through follow-up communication with the chief perfusionist under previous cases from this hospital, livanova (b)(4) learned that the water in the heater-cooler systems 3t in use is changed every day and they are stored dry.This is not in alignment with current instruction for use however reportedly the devices in use at the hospital are very clean and there is no sign of biofilm.The devices are located inside the operating theatre during use.The serial numbers of the device used for this surgery is unknown.However, two (2) over five (5) devices used at the hospital resulted to be contaminated.Only one serial number has been provided (b)(4) and a follow up report for the other cases associated to this specific device has been filed.The second serial number remains unknown as well as the type of contamination.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|