Model Number 12-80-10 |
Device Problem
Difficult to Open or Close (2921)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
There was no patient involvement.Livanova manufactures the electrical venous occluder (evo) the incident occurred in (b)(6).Livanova initiated an investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
Livanova (b)(4) received a report that a electrical venous occluder (evo) displayed an error code associated to the clamp closed position issue at the startup.There was no patient involvement.
|
|
Manufacturer Narrative
|
H.10: the unit was requested for investigation which confirmed the reported error code.The investigation results revealed that the linear actuator of the clamp was faulty and did not allow the clamp to close.The part will be replaced in order to solve the reported issue.
|
|
Event Description
|
See initial report.
|
|