DATEX-OHMEDA, INC. AISYS CS2 ANESTHESIA GAS MACHINE Back to Search Results
Device Problem Suction Failure (4039)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/15/2020
Event Type  malfunction  
Manufacturer Narrative
No report of patient involvement. The initial reporter is located outside the u. S. , and therefore this information is not provided due to country privacy laws. Unique identifier: (b)(4). A ge healthcare service representative performed a checkout of the system and confirmed the reported issue. The internal suction line was found to be kinked which was repaired and adjusted to resolve the reported issue.
 
Event Description
The hospital reported a malfunction resulting in the loss of suction. There was no report of patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameAISYS CS2
Type of DeviceANESTHESIA GAS MACHINE
Manufacturer (Section D)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison, WI 53718
Manufacturer Contact
john szalinski
MDR Report Key10986745
MDR Text Key221275731
Report Number2112667-2020-03452
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
PMA/PMN Number
K132530
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial
Report Date 12/10/2020
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received12/10/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received11/15/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/09/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage

-
-