This is being filed to report the blood clot requiring aspiration.
It was reported that this was a mitraclip procedure to treat functional mitral regurgitation (mr) with a grade of 4.
The steerable guide catheter (sgc) was advanced with resistance noted when passing through the septum.
After removing the dilator and guide wire, a bright object thought to be a blood clot was noted on echo on the tip of the sgc.
Aspiration was performed however the blood clot could not be removed.
The decision was made to continue the procedure.
The clip delivery system (cds) advanced to the left atrium (la) and the grippers tested.
The gripper on the anterior side lowered without issue however the gripper on the posterior side would not lower.
The lock lever was changed to the locked position and both grippers were able to lower; however the gripper lever latch came off.
The physician continued with use of the device and the clip implanted without further issues.
After removal of the cds, the sgc tip was confirmed on echo and the blood clot was gone.
The sgc was removed and the procedure completed with the mr reduced to 1.
There was no clinically significant delay in the procedure and no adverse patient effects.
No additional information was provided.
|