The device was not returned for evaluation.
A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.
The reported patient effect of dissection is listed in the xience prime everolimus eluting coronary stent systems instructions for use as a known patient effect of coronary stenting procedures.
A conclusive cause for the reported patient effect, and the relationship to the product, if any, cannot be determined.
However, the treatment appears to be related to the operational context of the procedure.
There is no indication of a product quality issue with respect to manufacture, design or labeling.
|
It was reported that the procedure performed was to treat a heavily tortuous, moderately calcified, left circumflex (lcx) artery.
A 2.
75x38mm xience prime was deployed and a distal dissection was noted.
A xience v was used to treat the dissection.
There was no adverse patient sequela and no clinically significant delay.
No additional information was provided.
|