Model Number 9438-05
Device Problem No Device Output (1435)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/02/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that transmitter failed error occurred. No product or data was provided for evaluation. Confirmation of the allegation and a probable cause could not be determined. No injury or medical intervention was reported.
 
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that transmitter failed error occurred. The product was evaluated. An external visual inspection was performed and passed. Voltage test was performed and failed due to 0 v. Data was received but does not reflect full investigation window. Confirmation of the allegation could not be determined. A probable cause could not be determined as the allegation was not found. No injury or medical intervention was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameDEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM
Type of DeviceCONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence dr.
san diego CA 92121
Manufacturer Contact
ashley spoto
6340 sequence dr.
san diego, CA 92121
8582000200
MDR Report Key10988218
MDR Text Key220963403
Report Number3004753838-2020-153767
Device Sequence Number1
Product Code MDS
Combination Product (y/n)N
Reporter Country CodeEZ
PMA/PMN Number
P120005
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Reporter Occupation
Type of Report Initial,Followup
Report Date 03/12/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received12/10/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number9438-05
Device Catalogue NumberSTT-GL-004
Device Lot Number5271284
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/14/2021
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received03/11/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/29/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-