Model Number 9438-05 |
Device Problem
No Device Output (1435)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 10/18/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).
|
|
Event Description
|
It was reported that transmitter failed error occurred.No product or data was provided for evaluation.Confirmation of the allegation and a probable cause could not be determined.No injury or medical intervention was reported.
|
|
Manufacturer Narrative
|
(b)(4).B5: describe event or problem - additional.D9: device availability ¿ additional.G3: date received by manufacturer - additional.H2: additional information/device evaluation.H3: device evaluated by manufacturer - additional.H6: event problem and evaluation codes - additional.
|
|
Event Description
|
It was reported that transmitter failed error occurred.The product was evaluated.An external visual inspection was performed and passed.Voltage check was performed and passed.Pairing test/download was performed and passed.A review of the share logs was performed and transmitter failed error was not found within the investigation window.The allegation was not confirmed.The probable cause could not be determined.No injury or medical intervention was reported.
|
|