RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problem
Therapeutic or Diagnostic Output Failure (3023)
|
Patient Problem
Insufficient Information (4580)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event: (b)(6) 2020, date of report: 10dec2020.
|
|
Event Description
|
It was reported the touchscreen was not sensitive.There was no patient involvement.The field service engineer (fse) confirmed the reported issue.The fse was unable to touch settings on the screen and determined the issue to be a screen failure.The fse replaced the touchscreen, performed calibration, and the issue was resolved.The device was returned to full functionality.
|
|