Model Number V3.0 |
Device Problems
Computer Operating System Problem (2898); Application Program Freezes, Becomes Nonfunctional (4031)
|
Patient Problem
Insufficient Information (4580)
|
Event Date 12/03/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
|
|
Event Description
|
A customer reported that before surgery began, the system froze.The system was rebooted to initiate the procedure.
|
|
Manufacturer Narrative
|
The previous report submitted for this event contained an error in h.1.- ¿summary report¿ was inadvertently selected.After a recent system update, a system error caused the inadvertent additional selection of ¿summary report¿ in h.1.The manufacturer internal reference number is: (b)(4).
|
|
Manufacturer Narrative
|
The company service representative examined the system and was able to confirm and replicate the reported event of a system freeze/lock-up.The nonconforming display printed circuit board (pcb) was replaced to address the issue.The system was then tested and met all product specifications.The system manufacturing device history record (dhr) was reviewed.Based on qa assessment, the product met specifications at the time of release.The root cause of the reported event can be attributed to the nonconforming display printed circuit board (pcb).The manufacturer internal reference number is: (b)(4).
|
|