Model Number V3.0
Device Problems Computer Operating System Problem (2898); Application Program Freezes, Becomes Nonfunctional (4031)
Patient Problem Insufficient Information (4580)
Event Date 12/03/2020
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
 
Event Description
A customer reported that before surgery began, the system froze.The system was rebooted to initiate the procedure.
 
Manufacturer Narrative
The previous report submitted for this event contained an error in h.1.- ¿summary report¿ was inadvertently selected.After a recent system update, a system error caused the inadvertent additional selection of ¿summary report¿ in h.1.The manufacturer internal reference number is: (b)(4).
 
Manufacturer Narrative
The company service representative examined the system and was able to confirm and replicate the reported event of a system freeze/lock-up.The nonconforming display printed circuit board (pcb) was replaced to address the issue.The system was then tested and met all product specifications.The system manufacturing device history record (dhr) was reviewed.Based on qa assessment, the product met specifications at the time of release.The root cause of the reported event can be attributed to the nonconforming display printed circuit board (pcb).The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFINITI VISION SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
MDR Report Key11002069
MDR Text Key221844870
Report Number2028159-2020-01107
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
PMA/PMN Number
K120912
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 03/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/14/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV3.0
Device Catalogue Number8065752160
Was Device Available for Evaluation? Yes
Date Manufacturer Received03/03/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-