C.R. BARD, INC. (BASD) -3006260740 M.R.I. IMPLANTABLE PORT, HICKMAN SINGLE-LUMEN, 9.6F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR Back to Search Results
Model Number 0602680
Device Problems Entrapment of Device (1212); Fracture (1260)
Patient Problem Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 11/13/2020
Event Type  Injury  
Manufacturer Narrative
As the lot number for the device was provided, a manufacturing review will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.(expiry date: 11/2024).
 
Event Description
It was reported that during port placement procedure, the guidewire of the device allegedly damaged and got stuck.Reportedly, the entire device was removed and another device was used in new access site.The patient status was normal.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
M.R.I. IMPLANTABLE PORT, HICKMAN SINGLE-LUMEN, 9.6F
Type of Device
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
judy ludwig
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key11002290
MDR Text Key221236586
Report Number3006260740-2020-20920
Device Sequence Number1
Product Code LJT
UDI-Device Identifier00801741025679
UDI-Public(01)00801741025679
Combination Product (y/n)N
Reporter Country CodeBR
PMA/PMN Number
K873213
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 11/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/14/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0602680
Device Catalogue Number0602680
Device Lot NumberREEQ1644
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/17/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-