• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TANDEM DIABETES CARE T:SLIM X2 INSULIN PUMP WITH BASAL-IQ TECHNOLOGY; AUTOMATED INSULIN DOSING, THRESHOLD SUSPEND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TANDEM DIABETES CARE T:SLIM X2 INSULIN PUMP WITH BASAL-IQ TECHNOLOGY; AUTOMATED INSULIN DOSING, THRESHOLD SUSPEND Back to Search Results
Model Number 1002717
Device Problem Wireless Communication Problem (3283)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/20/2020
Event Type  malfunction  
Manufacturer Narrative
Per the tandem user guide: ¿do not insert the sensor in sites other than the abdomen (belly) or upper buttocks (for ages 6¿17 only).Other sites have not been studied and are not approved.Use in other sites might cause sensor glucose readings to be inaccurate and could result in you missing severe hypoglycemia (low blood glucose) or hyperglycemia (high blood glucose) events.¿ the device is expected to be returned; however, the device has not yet been received.A supplemental report will be submitted if the device is received and evaluation is performed.
 
Event Description
It was reported that out of range issues occurred between the transmitter and pump greater than 15 minutes, with no continuous glucose monitor (cgm) sensor readings.Reportedly, the sensor was placed on the arm.Customer¿s blood glucose level was not impacted.Reportedly, the customer reverted to manual injections for insulin therapy.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
T:SLIM X2 INSULIN PUMP WITH BASAL-IQ TECHNOLOGY
Type of Device
AUTOMATED INSULIN DOSING, THRESHOLD SUSPEND
Manufacturer (Section D)
TANDEM DIABETES CARE
11075 roselle street
san diego CA 92121
Manufacturer Contact
mick trier
san diego, CA 92121
8584011451
MDR Report Key11011818
MDR Text Key221531388
Report Number3013756811-2020-143249
Device Sequence Number1
Product Code OZO
UDI-Device Identifier00853052007998
UDI-Public(01)00853052007998
Combination Product (y/n)N
PMA/PMN Number
K201214
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/15/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number1002717
Device Catalogue Number1004219
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/24/2021
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Date Manufacturer Received05/03/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/01/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-