• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CONCORD MANUFACTURING 2008T HEMODIALYSIS SYS., WITH CDX; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONCORD MANUFACTURING 2008T HEMODIALYSIS SYS., WITH CDX; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Model Number 190713
Device Problem Sparking (2595)
Patient Problem No Patient Involvement (2645)
Event Date 11/25/2020
Event Type  malfunction  
Manufacturer Narrative
Plant investigation: no parts were returned to the manufacturer for physical evaluation.Additionally, no on-site evaluation was performed by a fresenius field service technician (fst).A records review was performed on the reported serial number.An investigation of the device manufacturing records was conducted by the manufacturer.There were no non-conformances, or any associated rework identified during the manufacturing process which could be related to the reported event.In addition, the device history record (dhr) review confirmed the results of the in-progress and final quality control (qc) testing met all requirements.The investigation into the cause of the reported problem was not able to be confirmed.A definitive conclusion regarding the complaint incident cannot be reached without a physical examination of the complaint device.
 
Event Description
A user facility biomedical technician (biomed) reported sparking coming from a fresenius 2008t hemodialysis (hd) machine.The biomed was performing troubleshooting on the hd system which was pulled from service due to a low flow error and high conductivity alarm.During the biomed¿s initial inspection, they noticed the deaeration pump was not running.Therefore, the biomed started by rebuilding the deaeration pump.After completing the rebuild, the biomed plugged the pump into the distribution board and observed sparking coming from the board, where the deaeration pump pins connect.To isolate the issue, the biomed then replaced the deaeration pump and the distribution board.Both parts were reportedly known, good parts.When the biomed proceeded to plug the deaeration pump into the distribution board, once again, they observed sparking.At this point, the biomed contacted technical support (ts) for further assistance.As advised by ts, the biomed replaced the actuator test board and actuator cable, in addition to the deaeration pump and distribution board (for a second time).Completing these repairs resolved the reported issue.According to the biomed, there was no burning smell, smoke, melting, or arcing.The biomed also confirmed there were no flames.The biomed did not know how many hours were on the machine.No damage was identified on any other components.The machine was plugged into a hospital grade ground-fault circuit interrupter (gfci) outlet, and there was no previous history of the machine failing the electrical leakage test.The biomed confirmed the hd machine was returned to service.The faulty parts were discarded; no samples were available to be returned for evaluation.There was no patient involvement associated with the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2008T HEMODIALYSIS SYS., WITH CDX
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
CONCORD MANUFACTURING
director, quality systems
4040 nelson avenue
concord CA 94520
Manufacturer (Section G)
CONCORD MANUFACTURING
director, quality systems
4040 nelson avenue
concord CA 94520
Manufacturer Contact
matthew amaral
920 winter st
waltham, MA 02451
7816999758
MDR Report Key11012816
MDR Text Key221785332
Report Number2937457-2020-02254
Device Sequence Number1
Product Code KDI
UDI-Device Identifier00840861100897
UDI-Public00840861100897
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K093902
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 12/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/15/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number190713
Device Catalogue Number190713
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Device AgeMO
Date Manufacturer Received11/25/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/31/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-