OLYMPUS MEDICAL SYSTEMS CORP. VIDEOSCOPE "ENF-VQ"; RHINO-LARYNGO VIDEOSCOPE Back to Search Results
Model Number ENF-VQ
Device Problem Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
The subject device was not returned to omsc for evaluation.The exact cause of the reported event could not be conclusively determined at this time.If additional information becomes available, this report will be supplemented.
 
Event Description
Olympus medical systems corp.(omsc) was informed that during the incoming inspection for repair at the service department of olympus (b)(4)b(oekg), it was found that the component at the distal end of the subject device had fallen off.There was no report of patient injury associated with this event.
 
Manufacturer Narrative
This supplemental report is being submitted to provide additional information.The subject device was not returned to olympus medical systems corp.(omsc) for evaluation.Omsc reviewed the manufacturing history (dhr) of the subject device and confirmed no irregularity.The exact cause of the reported phenomenon could not be conclusively determined.However, based upon the information from olympus italia, there was the possibility that the reported phenomenon was attributed to the user handling, as there was a trace of physical damage on the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIDEOSCOPE "ENF-VQ"
Type of Device
RHINO-LARYNGO VIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key11018567
MDR Text Key227166752
Report Number8010047-2020-10467
Device Sequence Number1
Product Code EOB
Combination Product (y/n)N
PMA/PMN Number
K061313
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 01/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/16/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberENF-VQ
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received01/19/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-