There was no patient involvement.Livanova (b)(4) manufactures the electrical venous occluder (evo).The incident occurred in (b)(6).The unit was requested back to the manufacturer site for further investigation.The issue could be reproduced and two more error codes were displayed.The mechanical tolerances at the bearing seat of the linear actuator were found faulty and the centering of the plunger was not sufficient.Several part were need replacement in order to solve the problem.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|