MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION PRISMAFLEX M150 DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION PRISMAFLEX M150 DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 109990
Device Problem Fluid Leak (1250)
Patient Problem No Patient Involvement (2645)
Event Date 11/20/2020
Event Type  malfunction  
Manufacturer Narrative
Facility name: (b)(6). A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot. The device was not received for evaluation; therefore, a device analysis could not be completed. However, photos of the impacted set were provided. Their visual inspection observed an external leak fluid from the effluent port. The reported condition was verified. The reported fluid leak is due to an internal crack on the dialysate port. This kind of damage is typical of mechanical shock applied on the product leading to its breakage during transport. Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that an external fluid leak was observed at the waste connection port on a prismaflex m150 set during priming. There was no patient involvement. No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NamePRISMAFLEX M150
Type of DeviceDIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - MEYZIEU
7, av lionel terray, b.p. 126
meyzieu cedex rhone 69883
FR 69883
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key11020281
MDR Text Key225335288
Report Number8010182-2020-00297
Device Sequence Number1
Product Code KDI
Combination Product (y/n)Y
Reporter Country CodeTU
PMA/PMN Number
K080519
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 12/16/2020
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received12/16/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue Number109990
Device Lot Number20C0504
Was Device Available for Evaluation? No
Is the Reporter a Health Professional?
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received11/21/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unkown

-
-