Model Number 7655405
Device Problem Detachment of Device or Device Component (2907)
Patient Problems No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/23/2020
Event Type  malfunction  
Manufacturer Narrative
The information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The manufacturer has received the sample and is pending evaluation.Results are expected soon.A lot history review (lhr) of recu2723 showed no other similar product complaint(s) from this lot number.
 
Event Description
It was reported that the connector detached and fluid leaks occurred after reconnection.No other information was provided.
 
Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, complaint and batch history, applicable previous investigation(s), labeling, applicable manufacturing records, sample analysis and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of a hub detachment is confirmed but the exact cause remains unknown.One photo sample of a 4 fr groshong nxt connector was provided for evaluation.The two-piece connector is shown to be assembled with the catheter.The catheter appears to have been trimmed.The red hub is detached from the white compression sleeve and extension tubing.The characteristics of the components could not be clearly inspected from the image provided.Based on the description of the reported event and photo provided, possible contributing factors include tensile damage and excessive manipulation of the connector.Since the hub was shown to be detached, the complaint is confirmed but the exact contributing factors remain unknown.H3 other text : evaluation findings are in section h.11.
 
Event Description
It was reported that the connector detached and fluid leaks occurred after reconnection.No other information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GROSHONG NXT CLEARVUE 4F SL
Type of Device
CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
MDR Report Key11023782
MDR Text Key222321583
Report Number3006260740-2020-20959
Device Sequence Number1
Product Code LJS
UDI-Device Identifier00801741035272
UDI-Public(01)00801741035272
Combination Product (y/n)N
PMA/PMN Number
K034020
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,other
Type of Report Initial,Followup
Report Date 03/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/16/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number7655405
Device Catalogue Number7655405
Device Lot NumberRECU2723
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/25/2020
Date Manufacturer Received03/22/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-