Unique device identifier (udi) : (b)(4).The hakim valve was returned for evaluation.Device history record (dhr) - lot 4171051, conformed to the specifications when released to stock.Failure analysis - the valve was visually inspected; biological debris was noted inside valve and on the outside of the valve, biological debris was also noted inside of catheters.The position of the cam when valve was received was 70mmh2o.The valve was hydrated.The valve was tested for programming with programmer 82-3126 with serial number (b)(6) and programmer 82-3190 with serial number (b)(6), the valve failed the test, the cam mechanism did not move during the programming process.The valve was flushed and failed.The catheters were irrigated, failed the catheters were occluded with biological debris.The valve could not be reflux due to occlusion.The valve was dismantled and was examined under microscope at appropriate magnification and biological debris was found covering the valve mechanism and inside the catheters.The cam magnets were controlled and passed.The root cause for the issue reported by the customer is due to biological debris found covering the valve mechanism and inside the catheters.
|