• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC STEALTHSTATION S8 SYSTEM; INSTRUMENT, STEREOTAXIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC STEALTHSTATION S8 SYSTEM; INSTRUMENT, STEREOTAXIC Back to Search Results
Model Number 9735665
Device Problem Imprecision (1307)
Patient Problem Unspecified Tissue Injury (4559)
Event Date 12/10/2020
Event Type  malfunction  
Manufacturer Narrative
Other relevant device(s) are: product id: 9735740, serial/lot : n/a, version: (b)(4).A medtronic representative went to the site to perform a system check out and they found the system functioned as intended.The accuracy was tested with a blue spine model.One spin was done on either side of the imaging system.Both spins were found to be accurate.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information regarding a navigation system being used during a spinal procedure.It was reported that the surgeon alleged that the accuracy was off.The surgeon navigated to known anatomical landmarks prior to using the drill.However, the pilot holes that were drilled were off.The surgeon decided to re-spin, but the same thing happened.The surgeon decided to abort the use of navigation and continue the case with an imaging system.The site was using the right side tracker.There was no patient harm and the procedure was delayed by 20 minutes.
 
Event Description
Additional information received from a manufacturer representative reported that the amount of inaccuracy was unknown.
 
Manufacturer Narrative
See b5.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
H3, h6: the software investigation was unable to determine probable cause.Logs were reviewed but provided no additional insight regarding the cause of reported complaint.Without exam archives there is insufficient information to determine whether a software anomaly contributed to the reported behavior.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION S8 SYSTEM
Type of Device
INSTRUMENT, STEREOTAXIC
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
MDR Report Key11027358
MDR Text Key222037203
Report Number1723170-2020-03307
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00643169722187
UDI-Public00643169722187
Combination Product (y/n)N
PMA/PMN Number
K162309
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 03/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/17/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9735665
Device Catalogue Number9735665
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/05/2021
Date Manufacturer Received03/16/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
"SEE H10...."
Patient Age64 YR
Patient Weight57
-
-