Model Number PED2-450-16
Device Problem Activation Failure (3270)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/15/2020
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received a report that the when the pipeline shield was deployed, the distal end did not open fully, and the flow diverter remained partly crimped. The proximal part of the stent was also not fully apposed to the vessel wall. Balloon was used to post dilate. Stent apposed the vessel wall after post dilatation. A second pipeline shield was used inside the first one. There was no harm or injury to the patient, no actions required as a result of the event, and the patient was noted to be stable and recovering from the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NamePIPELINE FLEX W/SHIELD TECHNOLOGY
Type of DeviceINTRACRANIAL ANEURYSM FLOW DIVERTER
Manufacturer (Section D)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer (Section G)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer Contact
tricha miles
9775 toledo way
irvine, CA 92618
7635140379
MDR Report Key11037595
MDR Text Key225565973
Report Number2029214-2020-01341
Device Sequence Number1
Product Code OUT
Combination Product (y/n)N
Reporter Country CodeTC
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation
Type of Report Initial
Report Date 12/18/2020
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received12/18/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberPED2-450-16
Device Catalogue NumberPED2-450-16
Device Lot NumberB087995
Was Device Available for Evaluation? No
Is the Reporter a Health Professional?
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received12/15/2020
Was Device Evaluated by Manufacturer? No Answer Provided
Date Device Manufactured09/17/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-