• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION INVIVE; PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION INVIVE; PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P) Back to Search Results
Model Number V173
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem Muscle Stimulation (1412)
Event Date 10/25/2020
Event Type  Injury  
Event Description
It was reported that this patient presented to the emergency department with phrenic nerve stimulation, due to pacing resulting from this cardiac resynchronization therapy pacemaker (crt-p) triggering safety mode.Subsequently, the patient was admitted to the hospital.It was noted the patient was already in hospice care and the decision was made to not explant the device and to cancel all future device follow-up care.No other device programming options were available.The device remains in service.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INVIVE
Type of Device
PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key11042502
MDR Text Key222625110
Report Number2124215-2020-25571
Device Sequence Number1
Product Code NKE
UDI-Device Identifier00802526536632
UDI-Public00802526536632
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030005/S079
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 12/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/18/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date12/08/2013
Device Model NumberV173
Device Catalogue NumberV173
Device Lot Number100005
Was Device Available for Evaluation? No
Date Manufacturer Received10/25/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/21/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
Patient Age96 YR
-
-