It was determined during an onsite visit from a field service engineer (fse), the system completed daily calibration, and qa check, followed by four completed patient scans with no fault.On the fifth scan the system stopped, as reported by ct techs.The system completed a sixth scan with no fault, passed another daily calibration, and qa check.Per communication received from fse, an upgrade to the software version corrected the error the device experienced.It was found that even though the patient experienced an additional scan, the dose amount received is not considered to be harmful to the patient.
|