Model Number H1-S
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/18/2020
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Physician intended to use a hawkone atherectomy device with a 6fr sheath and non-medtronic guidewire during treatment of a 20mm plaque lesion in the patient¿s distal right anterior tibial artery. Slight vessel tortuosity and moderate calcification are reported. Artery diameter reported as 3m. Lesion exhibited 80% stenosis. Ifu was followed. Tip detachment is reported during retrieval of the device. The tip is reported to have separated at the hinge pin. The detached tip is reported to have become stuck in the sheath. No components remained in the patient. The physician was able to remove the detached portion as it remained attached to the wire inside the sheath. Access could not be maintained so post op imaging could not be completed. No patient injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameHAWKONE 6FR
Type of DeviceCATHETER, PERIPHERAL, ATHERECTOMY
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
EI
Manufacturer (Section G)
MEDTRONIC IRELAND
parkmore business park west
galway
EI
Manufacturer Contact
toni o'doherty
parkmore business park west
galway 
EI  
091708734
MDR Report Key11045285
MDR Text Key223614871
Report Number9612164-2020-04997
Device Sequence Number1
Product Code MCW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K161361
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation
Remedial Action Recall
Type of Report Initial,Followup,Followup,Followup
Report Date 02/16/2022
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received12/21/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberH1-S
Device Catalogue NumberH1-S
Device Lot Number0010297604
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/19/2022
Was Device Evaluated by Manufacturer? No Answer Provided
Date Device Manufactured07/02/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-0451-2022

-
-