Model Number 2545-00-026 |
Device Problem
Break (1069)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/09/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
Some pieces and devices from attune instruments have color sizes recorded.From one of the pieces, size 6 was released.There is a danger of this "film" with the number being in the patient´s joint.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).Investigation summary - the device associated with the reported event was not returned.Visual examination of the provided photograph found that the white lettering on the instrument was worn.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot - the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).Where the lot code was provided, a manufacturing records evaluation (mre) was not performed.
|
|