OGDEN MANUFACTURING PLANT OPTIFLUX 180NRE DIALYZER FINISHED ASSY.; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
|
Back to Search Results |
|
Model Number 0500318E |
Device Problem
Fluid/Blood Leak (1250)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 12/11/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Plant investigation: the plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
|
|
Event Description
|
A user facility clinical manager reported that a dialyzer blood leak occurred immediately after the initiation of the patient¿s hemodialysis (hd) treatment.The blood leak was noted as being an internal blood leak that could be visually observed within the arterial end of the dialyzer.The machine, a fresenius 2008t machine, alarmed appropriately with a blood leak alarm.Blood leak test strips were used and tested positive for the presence of blood.There was no defect, or damage seen on the dialyzer.The patient¿s estimated blood loss (ebl) was approximately 75 ml.There was no patient injury, adverse events, or medical intervention required as a result of this event.The patient was restarted on a new machine and treatment completed successfully with new supplies.The complaint device was reported to be unavailable to be returned to the manufacturer for evaluation.
|
|
Manufacturer Narrative
|
Plant investigation: the reported complaint was not confirmed as the complaint device was not returned for manufacturer evaluation.A production records review was performed on the reported lot.An investigation of the device history records (dhr) was conducted by the manufacturer.There was no indication of product non-acceptance or deviation in the manufacturing process related to the complaint event.This includes non-conformances, rework, labeling, process controls, and any other occurrence in production that was potentially related to the complaint.The lot met all release criteria.A definitive conclusion regarding the complaint incident cannot be reached without physical examination of the actual device.Therefore, the complaint is not confirmed.
|
|
Search Alerts/Recalls
|
|
|