Model Number 3500CP-G
Device Problem Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/01/2020
Event Type  malfunction  
Manufacturer Narrative
This report is based solely on medsun report (b)(4).Additional information was requested from the customer and has not been received.A dhr review could not be performed since there was no serial number provided for this device.At this time there is no evidence that a manufacturing non-conformity contributed to the reported complaint, and the instructions for use were reviewed and determined to provide adequate instructions and warnings for the safe and effective use of the device.Therefore, no corrective or preventative actions are necessary.All complaints are trended and reviewed by management on a monthly basis.As part of this monthly review, any excursion above the control limits for this failure mode will be assessed, documented and acted upon as warranted.Manufacturers record file no.(b)(4).
 
Event Description
Customer reported when the sweep gas was assessed, the knob on the liter side of sweep gas appeared to be off as the ball used to gauge the level of the sweep gas was not elevated.However, the knob, when turned to the left, was able to be tightened down and truly turned off.It is unclear where the issue was discovered and no patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SECHRIST AIR/OXYGEN GAS BLENDER, MODEL 3500CP-G LOW FLOW
Type of Device
GAS CONTROL UNIT, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
SECHRIST INDUSTRIES, INC
4225 e. la palma ave.
anaheim CA 92807
Manufacturer (Section G)
SECHRIST INDUSTRIES, INC
4225 e. la palma ave.
anaheim CA 92807
Manufacturer Contact
majid mashayekh
4225 e. la palma ave.
anaheim, CA 92807
7145798309
MDR Report Key11051878
MDR Text Key223270559
Report Number2020676-2020-00030
Device Sequence Number1
Product Code DTX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K023745
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/21/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3500CP-G
Device Catalogue Number3500CP-G
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/04/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-