Model Number ROSA ONE 3.1
Device Problems Computer Software Problem (1112); Mechanical Problem (1384)
Patient Problems No Code Available (3191); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/07/2020
Event Type  malfunction  
Manufacturer Narrative
The device has not been evaluated yet for investigation purpose. Once the evaluation is performed, a follow-up medwatch report will be submitted. Unique identifier (udi) #: (b)(4).
 
Event Description
Before the registration, a restart did not work, the clinical representative (cr) had to powercycle the entire machine. Registration then went smoothly and everything accurate. The cr and the surgeon then navigated to the first trajectory. After placing the bolt the surgeon moved the robot away from the patient in the axial plane to which it appeared to be a safe distance. Upon moving back to home position the robot collided with the patient, they immediately stopped movement, at that point the software became unresponsive and the robot arm would not move away from the patient¿s head. The cr did a restart, it was still unresponsive so he then powercycled the entire machine again. After this, they were then able to move the arm away from the patient¿s head in the axial plan, move it to a far safe distance and return it to home position. They then put it in the park position so they could acquire an intraoperative ct with the brianlab airo. They did resume the case and they were able to finish the implant.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameROSA ONE
Type of DeviceCOMPUTER-ASSISTED SURGICAL DEVICE
Manufacturer (Section D)
MEDTECH SA
zac eureka
900 rue du mas de verchant
montpellier, languedoc-roussillon 34000
FR 34000
Manufacturer (Section G)
MEDTECH SA
zac eureka
900 rue du mas de verchant
montpellier, languedoc-roussillon 34000
FR 34000
Manufacturer Contact
jay sharma
1520 tradeport drive
jacksonville, FL 32218
9047414400
MDR Report Key11054942
MDR Text Key227549247
Report Number3009185973-2020-00325
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K200511
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date 03/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/22/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberROSA ONE 3.1
Device Catalogue NumberROSAS00203
Device Lot Number3.1.4.1650
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional?
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received03/12/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/03/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A

Patient Treatment Data
Date Received: 12/22/2020 Patient Sequence Number: 1
-
-