The system was implanted on friday, (b)(6) 2020, and from the start there was the problem that the internal pressure was not displayed and instead only a dashed line could be seen in the display.
As a result, the deltap display was also missing.
All measures to remedy the problem did not lead to success.
The connector of the pressure sensor has been disconnected from the console and the disposable product.
The system was shut down and restarted.
The internal pressure was not displayed.
The patient was at risk.
The remaining pressures were zeroed several times to the atmosphere, and did not appear to be valid.
The complete system was changed on (b)(6) 2020.
Since then everything has worked perfectly.
A defect in the console can be ruled out.
The set is no longer available; it has already been discarded complaint id: (b)(4).
|