Device evaluation determined that the reported issue was not able to be duplicated.
The device was tested and all functions were found to be within standard specifications, no issues were observed.
Further testing found that the unit worked with the reference generator with no issues or abnormalities noted.
In addition, the unit was tested with the customer's returned generator ((b)(4)), and the unit worked properly with no issue observed.
Additionally, burn in testing was performed, and the unit passed testing with no issues observed.
Based on evaluation findings the reported issue was not confirmed as the device passed all testing with no issues, or abnormalities observed.
|