GYRUS ACMI INC CW-USLTA CYBERWAND ADVANCED TRANSDUCER LITHOTRIPTOR, ELECTRO-HYDRAULIC Back to Search Results
Model Number CW-USLTA
Device Problems No Device Output (1435); Output Problem (3005)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Device evaluation determined that the reported issue was not able to be duplicated. The suction port was found with a big dent, however, no other functional issues were observed. The device was tested and all functions were found to be within standard specifications, no issues were observed. Further testing found that the unit worked with the reference generator with no issues or abnormalities noted. In addition, the unit was tested with the customer returned generator ((b)(4)), and the unit worked properly with no issue observed. Additionally, burn in testing was performed and the unit passed testing with no issues observed. Based on evaluation findings the reported issue was not confirmed as the device passed all testing with no issues, or abnormalities observed.
 
Event Description
It was reported that during an unspecified procedure the device transducer motor stopped working after being in use for 5 minutes. The user power cycled the device for a few minutes. The intended procedure was not completed. There were no other details provided regarding the event. There was no patient harm or injury reported. No user injury reported. This report is related to reports with patient identifiers: (b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameCW-USLTA CYBERWAND ADVANCED TRANSDUCER
Type of DeviceLITHOTRIPTOR, ELECTRO-HYDRAULIC
Manufacturer (Section D)
GYRUS ACMI INC
136 turnpike road
southborough PA 01772
Manufacturer Contact
kenneth pittman
9600 louisiana avenue north
brooklyn park, MA 55445
9013785969
MDR Report Key11055447
MDR Text Key223164061
Report Number3011050570-2020-00189
Device Sequence Number1
Product Code FFK
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K120303
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation
Type of Report Initial,Followup,Followup
Report Date 03/31/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received12/22/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberCW-USLTA
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/25/2020
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received03/30/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-