Catalog Number UNKNOWN
Device Problems Filling Problem (1233); Calibration Problem (2890); Gas/Air Leak (2946)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/04/2020
Event Type  malfunction  
Manufacturer Narrative
Complete initial reporter name: (b)(6).The device has not been returned to the manufacturer so we are unable to complete an evaluation.If provided we will send a supplemental report with our additional findings.Complaint record id # (b)(4).
 
Event Description
It was reported that during intra-aortic balloon (iab) therapy, a helium leak from the iab occurred.The iab then failed to auto-fill and calibrate.There was no patient harm or adverse event reported.
 
Event Description
It was reported that during intra-aortic balloon (iab) therapy, a helium leak from the iab occurred.The iab then failed to auto-fill and calibrate.There was no patient harm or adverse event reported.
 
Manufacturer Narrative
The device has not been returned to the manufacturer so we are unable to complete an evaluation.If provided we will send a supplemental report with our additional findings.The failure modes are addressed in the risk file and are operating within its risk profile.The ifu addresses the reported failures.There were no ncmrs identified which could cause or contribute to the reported failure.The investigation does not indicate that the device was inadvertently released as non-conforming or an adulterated product or was a counterfeit.The complaint history review did not identify an adverse trend (increase in number of complaints over past three (3) months).Based on the rational provided above, no escalation to the capa process is required.Reference complaint #:(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN IAB
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - FAIRFIELD
15 law drive
fairfield NJ 07004
MDR Report Key11056774
MDR Text Key227381672
Report Number2248146-2020-00689
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 04/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/22/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNKNOWN
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Yes
Date Manufacturer Received04/26/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
CARDIOSAVE / (B)(4)
-
-