Model Number V200 |
Device Problems
Device Alarm System (1012); Tidal Volume Fluctuations (1634); Device Sensing Problem (2917)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event: (b)(6) 2020.Date of report: 22dec2020.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
|
|
Event Description
|
The customer reported high tidal volume and alarm indicator not blinking.There was no patient involvement.The manufacturer¿s international service technician confirmed the reported issue.The manufacturer¿s international service technician replaced the defective sensor, air/exhalation flow, flex cable, and sensor pcb to address the reported problem.The unit successfully passed the required performance verification test.
|
|